CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC.
March 2002
CyberAmp 380
OPERATOR'S MANUAL
Part Number 2500-117 Rev F, Printed in U.S.A.
Copyright 1992, 1996, 2002 Axon Instruments, Inc.
No part of this manual may be reproduced, stored in a retrieval system, or transmitted, in any form or
by any means, electronic, mechanical, photocopying, microfilming, recording, or otherwise, without
written permission from Axon Instruments, Inc.
QUESTIONS? Telephone +1(510) 675-6200 or Fax +1(510) 675-6300
Przeglądanie stron 0
1 2 3 4 5 6 ... 156 157

Podsumowanie treści

Strona 1 - CyberAmp 380

CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. March 2002 CyberAmp 380 OPERATOR'S MANUAL Part Number 2500-117 Rev F,

Strona 2

CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. SAFETY The CyberAmp 380 and associated probes are not intended to be used and should no

Strona 3 - Declaration of Conformity

Page 82 QUICK300 CONTROL SOFTWARE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Command Line Options A number of options can be se

Strona 4

CyberControl SOFTWARE Page 83 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. CyberControl SOFTWARE CyberControl allows the user wit

Strona 5 - IMPORTANT INFORMATION

Page 84 CyberControl SOFTWARE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Starting CyberControl When CyberControl is opened, the

Strona 6 - RENSEIGNMENTS IMPORTANTS

CyberControl SOFTWARE Page 85 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. the entire spreadsheet. With all of the cell selected

Strona 7 - WICHTIGE INFORMATIONEN

Page 86 CyberControl SOFTWARE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 5) PRE-FILTER AMPLIFICATION COLUMN PC: The Initial G

Strona 8 - INFORMACION IMPORTANTE

CyberControl SOFTWARE Page 87 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 9) OUTPUT AMPLIFICATION PC: The Total Gain field disp

Strona 9 - Explicación de símbolos

Page 88 CyberControl SOFTWARE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Zeroing Channels PC: The CyberAmp series of signal co

Strona 10 - SEGURIDAD

CyberControl SOFTWARE Page 89 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. SmartProbe probe is disconnected or changed, these cha

Strona 11 - VERIFICATION

Page 90 CyberControl SOFTWARE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 3) The Status Scanning Rate section controls the numb

Strona 12

CyberControl SOFTWARE Page 91 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Save Settings… saves the current settings of the selec

Strona 13 - TABLE OF CONTENTS

i CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. COPYRIGHT The circuits and information in this manual are copyrighted and must not

Strona 14

Page 92 CyberControl SOFTWARE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. The File Menu The file menu is used for opening and sa

Strona 15

CyberControl SOFTWARE Page 93 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. The Cyber Menu The Cyber menu provides access to the s

Strona 16

Page 94 PROGRAMMER'S GUIDE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. PROGRAMMER'S GUIDE This chapter provides inform

Strona 17

PROGRAMMER'S GUIDE Page 95 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. "AT" Command Protocols The "AT"

Strona 18

Page 96 PROGRAMMER'S GUIDE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Libraries and Example Code The contents of the CD di

Strona 19 - INTRODUCTION

PROGRAMMER'S GUIDE Page 97 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. CyberAmp 380 Command Set  Details In the descriptio

Strona 20 - Page 2 INTRODUCTION

Page 98 PROGRAMMER'S GUIDE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. "D"  DC offset in microvolts Format ATd

Strona 21 - INTRODUCTION Page 3

PROGRAMMER'S GUIDE Page 99 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. "E"  EEPROM read/write/verify/serial# F

Strona 22

Page 100 PROGRAMMER'S GUIDE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. "F"  Frequency of low-pass filter Form

Strona 23 - FUNCTIONAL DESCRIPTION

PROGRAMMER'S GUIDE Page 101 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. "G"  Gain Format ATdGnxggg where d =

Strona 25

Page 102 PROGRAMMER'S GUIDE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. "L" - Load factory defaults Format ATd

Strona 26 - General Features

PROGRAMMER'S GUIDE Page 103 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. "O"  Overload report and reset Format

Strona 27

Page 104 PROGRAMMER'S GUIDE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. "S"  Status report Format ATdSn where

Strona 28

PROGRAMMER'S GUIDE Page 105 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. "T"  Test oscillators Format ATdTO+ T

Strona 29 - FUNCTIONAL CHECKOUT

Page 106 PROGRAMMER'S GUIDE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. "W"  Write settings to main memory For

Strona 30

PROGRAMMER'S GUIDE Page 107 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. The following description of the library interface

Strona 31

Page 108 PROGRAMMER'S GUIDE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 2) integer SetDeviceNumber (integer DevNum) This p

Strona 32

PROGRAMMER'S GUIDE Page 109 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 9) integer SetFilter (integer Channel, integer Fil

Strona 33 - TYPICAL CONFIGURATIONS

Page 110 PROGRAMMER'S GUIDE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 15) integer GetChannelStatus (integer Channel, ref

Strona 34

PROGRAMMER'S GUIDE Page 111 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Example Programs The following BASIC example progra

Strona 35

iii CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. TABLE OF CONTENTS INTRODUCTION ...

Strona 36

Page 112 PROGRAMMER'S GUIDE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. The second example program displays the uninterpret

Strona 37 - Changing the Fuse

PROGRAMMER'S GUIDE Page 113 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. The following example program sets the gain of chan

Strona 38 - Grounding and Hum

Page 114 PROGRAMMER'S GUIDE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. The following example program sets the gain, filter

Strona 39 - Input Coupling

PROGRAMMER'S GUIDE Page 115 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. if C300.SetCoupling% (1, C300.NEGINPUT, C300.COUPL

Strona 40 - Expansion

Page 116 PROGRAMMER'S GUIDE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 5) If the investigator uses BNC cables to connect

Strona 41 - Low-Pass Filter

SPECIFICATIONS Page 117 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. SPECIFICATIONS NUMBER CHANNELS 8 ANALOG INPUTS Input signa

Strona 42 - Amplitude Characteristics

Page 118 SPECIFICATIONS CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. NOTCH FILTER Frequency: Tuneable from 45-70 Hz. Depth: 40 d

Strona 43 - 10-90% Rise Time

SPECIFICATIONS Page 119 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. OTHER FEATURES Electrode test: ±0.5 V square wave applied si

Strona 44 - 3 dB frequency of

Page 120 SPECIFICATIONS CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Click mode: Signal directly applied to speaker. Bandwidth:

Strona 45

SPECIFICATIONS Page 121 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. POWER AND DIMENSIONS Line voltage: 100-125 Vac or 200-250 Vac

Strona 46 - 100 1.6

iv iv CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. AC Coupling and Autozeroing ...

Strona 47 - Saturation

Page 122 SPECIFICATIONS CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 8 7 6 5 4 3 2 1 15 14 13 12 11 10 9 CHANNEL #3 OUTPUT CHAN

Strona 48 - Notch Filter

REFERENCES Page 123 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. REFERENCES References Cain, C. & Welch, A.J. (1974) Thin-f

Strona 50 - Audio Monitor

ADJUSTMENT PROCEDURES Page 125 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. ADJUSTMENT PROCEDURES There are very few adjustments

Strona 51 - External Input

Page 126 ADJUSTMENT PROCEDURES CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 2 V 10 ms Figure 39. The upper trace shows the ou

Strona 52 - Electrode Test

CIRCUIT DIAGRAMS Page 127 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. CIRCUIT DIAGRAMS REQUEST FORM All the information that you

Strona 54

SERIAL COMMUNICATION CABLES Page 129 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. SERIAL COMMUNICATIONS CABLES Most serial comm

Strona 56 - Active Probes

LIST OF FIGURES AND TABLES Page 131 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. LIST OF FIGURES AND TABLES Fig. 1 Safety symbol

Strona 57 - Common-Mode Rejection Ratio

v CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Self Heating...

Strona 59 - Troubleshooting

WARRANTY Page 133 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. WARRANTY Axon Instruments warrants its non-consumable hardware pr

Strona 61 - TRANSDUCERS

WARNING Page 135 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. WARNING Shipping the CyberAmp and AI 400 SmartProbes The CyberAmp i

Strona 63 - TRANSDUCERS Page 45

INDEX Page 137 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. A/D interface, 15 AC coupling, 26 Adjustment Procedures, 125 AI Series

Strona 65 - Metal Microelectrodes

INDEX CYBERAMP 380, COPYRIGHT MAY 1996, AXON INSTRUMENTS, INC.

Strona 66 - Page 48 TRANSDUCERS

vi vi CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Wheatstone Bridge Strain Gauge Circuits...

Strona 67 - Acceleration

vii CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Menus ...

Strona 69 - AI 400 SERIES SmartProbes

INTRODUCTION Page 1 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. INTRODUCTION The CyberAmp 380 provides the front-end signal cond

Strona 71 - Transducer Adapter

Page 2 INTRODUCTION CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Below is a partial list of transducers that can be accommodated

Strona 72 - Input Connector Color Scheme

INTRODUCTION Page 3 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. In many recording situations, sound is helpful in determining th

Strona 74 - Connector

FUNCTIONAL DESCRIPTION Page 5 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. FUNCTIONAL DESCRIPTION The CyberAmp 380 programmable

Strona 75 - Example Circuits

Page 6 FUNCTIONAL DESCRIPTION CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Figure 2. This figure shows the signal processi

Strona 76 - Basic Thermistor Circuit

FUNCTIONAL DESCRIPTION Page 7 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 4) Gain. The gain can be set at discrete values rang

Strona 77

Page 8 FUNCTIONAL DESCRIPTION CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 9) Input connectors. There are two inputs for each c

Strona 78 - Amplifier Probes

FUNCTIONAL DESCRIPTION Page 9 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 5) LINK connector. The outputs from all eight channe

Strona 79 - Photodiode Amplifier

Page 10 FUNCTIONAL DESCRIPTION CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 10) RS-232 interface. The CyberAmp communicates wit

Strona 80 - Rotary Potentiometers

FUNCTIONAL CHECKOUT Page 11 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. FUNCTIONAL CHECKOUT A new CyberAmp 380 should be subject

Strona 81 - Foot Switch

CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Declaration of Conformity Manufacturer: Axon Instruments, Inc. 3280 Whipple Road

Strona 82

Page 12 FUNCTIONAL CHECKOUT CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. The computer will respond with the following status info

Strona 83 - CMD300 CONTROL SOFTWARE

FUNCTIONAL CHECKOUT Page 13 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 9) Issue the command: CMD300 FILTER OFF GAIN 1 to bypas

Strona 85 - Interactive Mode

TYPICAL CONFIGURATIONS Page 15 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. TYPICAL CONFIGURATIONS The CyberAmp 380 excels when c

Strona 86 - Command-File Mode

Page 16 TYPICAL CONFIGURATIONS CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Almost any A/D interface can be used. The Digidata 1

Strona 87 - Command Reference

TYPICAL CONFIGURATIONS Page 17 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Figure 5. Here no interface box is us

Strona 88

Page 18 TYPICAL CONFIGURATIONS CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Figure 7. In this configuration, the c

Strona 89

GENERAL INFORMATION Page 19 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. REFERENCE SECTION — GENERAL INFORMATION Power Supply Vo

Strona 90

Page 20 GENERAL INFORMATION CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Grounding and Hum Grounding The analog ground inside th

Strona 91 - CMD300 OVERLOAD

GENERAL INFORMATION Page 21 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 8) Experiment. While hum can be explained in theory (

Strona 93

Page 22 GENERAL INFORMATION CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Expansion To increase the number of signal conditioning

Strona 94 - CMD300 REPEAT OVERLOAD

PRINCIPLES OF OPERATION Page 23 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. REFERENCE SECTION -- PRINCIPLES OF OPERATION Low-P

Strona 95

Page 24 PRINCIPLES OF OPERATION CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Amplitude Characteristics For signals at the selec

Strona 96

PRINCIPLES OF OPERATION Page 25 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Typically, as the order of a filter increases, the

Strona 97 - QUICK300 CONTROL SOFTWARE

Page 26 PRINCIPLES OF OPERATION CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Pre-Filter vs. Post-Filter Gain When low-level sig

Strona 98

PRINCIPLES OF OPERATION Page 27 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 500 ms 0.1 Hz 1 Hz 10 Hz INPUT Figure 10. A 1 Hz

Strona 99

Page 28 PRINCIPLES OF OPERATION CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. AUTOZERO AUTOZERODC WITH 0.1 Hz 1 Hz 10 Hz DC Fi

Strona 100 - Command Line Options

PRINCIPLES OF OPERATION Page 29 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Saturation The AC coupling circuit is the first ci

Strona 101 - CyberControl SOFTWARE

Page 30 PRINCIPLES OF OPERATION CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Pre-filter Input-referred Input-referred Gain

Strona 102 - Starting CyberControl

PRINCIPLES OF OPERATION Page 31 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. A B Figure 12. Part A shows an inappropriate use

Strona 103

IMPORTANT INFORMATION CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. WARNING IF THIS EQUIPMENT IS USED IN A MANNER NOT SPECIFIED B

Strona 104

Page 32 PRINCIPLES OF OPERATION CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Audio Monitor When monitoring data, experimenters

Strona 105 - Keyboard Commands

PRINCIPLES OF OPERATION Page 33 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. THRESHOLD SQUELCH THRESHOLD SQUELCH OV Figure 1

Strona 106

Page 34 PRINCIPLES OF OPERATION CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Electrode Test The electrode resistance is measure

Strona 107 - Communications Settings

PRINCIPLES OF OPERATION Page 35 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Figure 15. Change in the waveform of a

Strona 108 - PC Menus

Page 36 PRINCIPLES OF OPERATION CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Box 1 1) From Equation 2; RMVGain Ve=−100Ω 2) F

Strona 109 - Macintosh Menus

PRINCIPLES OF OPERATION Page 37 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Input Coupling To properly measure the electrode i

Strona 110

Page 38 PRINCIPLES OF OPERATION CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Active Probes Active SmartProbes have headstages i

Strona 111

PRINCIPLES OF OPERATION Page 39 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. resistor. Knowing the measured current (I), the r

Strona 112 - PROGRAMMER'S GUIDE

Page 40 PRINCIPLES OF OPERATION CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. The input resistance of the CyberAmp 380 main unit

Strona 113

PRINCIPLES OF OPERATION Page 41 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Troubleshooting 1) General It has been our experie

Strona 114 - Libraries and Example Code

RENSEIGNMENTS IMPORTANTS CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. ATTENTION L'EMPLOI DE CE MATERIEL D'UNE MANIERE DI

Strona 116 - > end of response

TRANSDUCERS Page 43 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. TRANSDUCERS Many different types of transducers are used in scie

Strona 117

Page 44 TRANSDUCERS CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. thermistor types. These can be custom interfaced to the CyberAm

Strona 118 - 2 4 6 8 10 ... 28000 30000

TRANSDUCERS Page 45 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. available thermocouples have diameters as small as 25 µm and the

Strona 119

Page 46 TRANSDUCERS CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. EMG, EEG, ECG Recording 1) EMG An electromyogram (EMG) can be re

Strona 120

TRANSDUCERS Page 47 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Nerve Cuffs Nerve cuff recordings have a frequency response up t

Strona 121 - AT8R06000

Page 48 TRANSDUCERS CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. capacitively couple into the EMG leads. This problem can be eli

Strona 122

TRANSDUCERS Page 49 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Pressure High quality pressure transducers have historically bee

Strona 123

Page 50 TRANSDUCERS CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Self Heating When an excitation current or voltage is applied ac

Strona 124 - D5=-1234500

AI 400 SERIES PROBES Page 51 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. AI 400 SERIES SmartProbes This section describes the cu

Strona 125 - Support Routine Details

WICHTIGE INFORMATIONEN CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. UNZULÄSSIGE VERWENDUNG DIESER APPARAT IST NICHT VORGESEHEN,

Strona 126

Page 52 AI 400 SERIES PROBES CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. AI 401, x10, Low-Noise Differential Amplifier This ampl

Strona 127

AI 400 SERIES PROBES Page 53 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. The AI 403 picoammeter probe is applicable to a broad r

Strona 128

Page 54 AI 400 SERIES PROBES CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Kits AI 490 Connector Kit The AI 490 connector kit help

Strona 129 - Example Programs

AI 400 SERIES PROBES Page 55 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. When a SmartProbe is added to the system, it is common

Strona 130

Page 56 MAKING YOUR OWN ADAPTER OR ACTIVE PROBE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. MAKING YOUR OWN ADAPTER OR ACTIVE PR

Strona 131

MAKING YOUR OWN ADAPTER OR ACTIVE PROBE Page 57 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. EEPROM A 256-byte serial EEPROM (i.e

Strona 132

Page 58 MAKING YOUR OWN ADAPTER OR ACTIVE PROBE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Series 700 Thermilinear Circuit T 1

Strona 133

MAKING YOUR OWN ADAPTER OR ACTIVE PROBE Page 59 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. software analysis. AD590 Integrated

Strona 134

Page 60 MAKING YOUR OWN ADAPTER OR ACTIVE PROBE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Type J and Type K Thermocouples AN

Strona 135 - SPECIFICATIONS

MAKING YOUR OWN ADAPTER OR ACTIVE PROBE Page 61 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Current Probes Photodiode Amplifier

Strona 136 - Page 118 SPECIFICATIONS

INFORMACION IMPORTANTE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. ADVERTENCIA SI ESTE EQUIPO SE USA DE MANERA NO ESPECIFICADA PO

Strona 137 - Offset range: ±10 V

Page 62 MAKING YOUR OWN ADAPTER OR ACTIVE PROBE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 5.000 V ( 4 ) INPUT ( 1 ) INPUT ( 2

Strona 138 - Page 120 SPECIFICATIONS

MAKING YOUR OWN ADAPTER OR ACTIVE PROBE Page 63 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Panel Switch Foot Switch C2 C1 R1 1

Strona 140 - Page 122 SPECIFICATIONS

CMD300 CONTROL SOFTWARE Page 65 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. CMD300 CONTROL SOFTWARE For PCs, the CyberAmp 380 is

Strona 141 - REFERENCES

Page 66 CMD300 CONTROL SOFTWARE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. respecified unless the group is changed. The syntax

Strona 142

CMD300 CONTROL SOFTWARE Page 67 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. CMD300 STATUS DEFAULTS Produces: Program CMD300 Ve

Strona 143 - ADJUSTMENT PROCEDURES

Page 68 CMD300 CONTROL SOFTWARE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Entering an alphanumeric key as the first character

Strona 144 - Other Adjustments

CMD300 CONTROL SOFTWARE Page 69 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Command Reference Summary of Commands The CyberAmp c

Strona 145 - CIRCUIT DIAGRAMS Page 127

Page 70 CMD300 CONTROL SOFTWARE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Command Details of Basic Commands The following basi

Strona 146

CMD300 CONTROL SOFTWARE Page 71 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 4) COUPLING + <coupling> or COUPLING

Strona 147 - SERIAL COMMUNICATIONS CABLES

CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Explanation of symbols Explication des symboles Erklärung der verwendeten symbole

Strona 148

Page 72 CMD300 CONTROL SOFTWARE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 7) FINDSPEED This command determines the optimal spe

Strona 149 - LIST OF FIGURES AND TABLES

CMD300 CONTROL SOFTWARE Page 73 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 12) OFFSET <offset value> The OFFSET comma

Strona 150

Page 74 CMD300 CONTROL SOFTWARE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. The information indicates that for channel number 1

Strona 151 - WARRANTY

CMD300 CONTROL SOFTWARE Page 75 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Command Details of Advanced Commands The following a

Strona 152

Page 76 CMD300 CONTROL SOFTWARE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 27) RAW The RAW command allows the user to pass an a

Strona 153 - WARNING

CMD300 CONTROL SOFTWARE Page 77 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 33) STEP This command sets the program into step-by-

Strona 154

Page 78 CMD300 CONTROL SOFTWARE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. 36) WRITEPROBE <probe number> <prob

Strona 155 - INDEX Page 137

QUICK300 CONTROL SOFTWARE Page 79 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. QUICK300 CONTROL SOFTWARE The Quick300 spreadsheet

Strona 156

Page 80 QUICK300 CONTROL SOFTWARE CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. port speed should be set to fast unless a communic

Strona 157

QUICK300 CONTROL SOFTWARE Page 81 CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC. Positive input, Negative input. The positive and